NUKLEONIKA 2009, 54(3):193-198
A LABORATORY FACILITY (RTD) TO STUDY RADON TRANSPORT THROUGH MODELED SOIL BED: RESULTS OF PRELIMINARY MEASUREMENTS
Krzysztof Kozak, Jadwiga Mazur, Hayk Hovhannisyan
Laboratory of Radiometric Expertise, The Henryk Niewodniczański Institute of Nuclear
Physics, Polish Academy of Sciences,
152 Radzikowskiego Str., 31-342 Kraków, Poland
The paper describes a laboratory facility (radon transport device - RTD) which makes it possible to study radon transport. The measuring position is a vertical cylindrical vessel with a height of 202 cm and a diameter of 24 cm. It can be filled with sand, gravel or other soil materials to be studied. The facility is providing radon gas from the source to the studied material and making it possible to measure radon concentration at different vertical distances from the source. The parameters of the medium (temperature, humidity) can be measured at the same time. The preliminary measurements using the RTD with sand as medium are presented.
Close X |